FAQ
ChIP Troubleshooting Guide 3
-
Check the possible causes and find out the solution!
Possible Causes
What You Can Do?
►
Not enough cells/chromatin
Add enough chromatin for each IP experiment. We suggest using at least 25 μg of chromatin for each IP.
►
Incorrect Protein A/G/L used
Make sure that the Protein A/G/L beads are capable of binding to the antibody subclass being used.
►
Cross-linking process too long
Over Cross-linking with formaldehyde might mask the antibody binding sites and reduce antibody binding ability. It is advisable to optimize the cross-linking steps by using different concentration of formaldehyde or reducing cross-linking time.
►
Not enough antibody
Titre antibody amount used for each IP to determine the optimal condition. Up to 10ug of antibody can be used for each IP experiment.
►
Washes too stringent
Reduce the number of washes. Reduce salt concentration in the wash buffer.
►
Antibody not capable of
immunoprecipitationTry a different antibody. Try polyclonal antibody if monoclonal antibody does not work well.
►
The chromatin size might be too small
Make sure that the shearing condition is not too harsh which might results in fragments of DNA smaller than what the primers are able to amplify.
►
Incomplete elution from the Protein
A/G/L beadsIncubate beads in elution buffer at 65°C with frequent mixing.
-
Check the possible causes and find out the solution!
Possible Causes
What You Can Do?
►
PCR reagent might be contaminated
Prepare new solutions from stock.
-
Check the possible causes and find out the solution!
Possible Causes
What You Can Do?
►
Inadequate washing
Use a more stringent washing buffer. Try to use a high salt washing buffer or increase the number of washes.
►
Non specific binding to Protein A,G or L
Include a pre-clear step by incubating lysate with Protein A/G/L agarose beads.
►
Too much DNA template added
to the PCR reaction, or too many
cycles of amplificationAdd less DNA template or reduce the number of cycles of amplification. Alternatively, real-time PCR can be used for the detection of ChIPed DNA products.
►
Buffers may be contaminated
Use freshly prepared lysis or wash buffers.
IP Troubleshooting Guide 3
-
Check the possible causes and find out the solution!
Possible Causes
What You Can Do?
►
Insufficient antibody
Check the recommended amount of antibody as indicated in the datasheet. Titrate and optimize the optimal antibody amount used per IP experiment.
►
Washes too stringent
Reduce the number of washes. Reduce salt concentration in the wash buffer.
►
Incorrect Protein A/G/L used
Make sure that the Protein A/G/L beads are capable of binding to the antibody subclass being used.
►
Target protein not present in the
sample usedMake sure that the target protein is expressed at a relatively high level in the sample used by including an Input sample in the WB.
►
Incorrect Lysis buffer used
Make sure that the lysis buffer used is not over-denaturing and destroy the native conformation of the target proteins.
►
Antibody not capable of
immunoprecipitationTry a different antibody. Try polyclonal antibody if monoclonal antibody does not work well.
-
Check the possible causes and find out the solution!
Possible Causes
What You Can Do?
►
Secondary antibody rcognizes heavy / light chain denatured from primary
antibodyUse secondary antibodies which only recognizes native form of IgG for immunoblotting.
-
Check the possible causes and find out the solution!
Possible Causes
What You Can Do?
►
Inadequate washing
Use a more stringent washing buffer. Try to use a high salt washing buffer or add 0.2% SDS or 1% Tween20 to washing buffer. Increase the number of washes.
►
Concentration of antibodies too high
Check the recommended amount of antibody as indicated in the datasheet. Titrate and optimize the optimal antibody amount used per IP experiment.
►
Non specific binding to Protein A,G or L
Pre-block beads with BSA. Incubate beads with 2%BSA in PBS for 1 hour wash in PBS before use.
►
Non specific binding to agarose beads
Include a pre-clear step by incubating lysate with Protein A/G/L agarose beads.
►
Antibody not specific enough
Use affinity purified and pre-absored antibody for IP experiments.
►
Sample degradation
Add adequate protease inhibitors and phosphatase inhibitors throughout sample preparation and Ip steps.
Western Blot Troubleshooting Guide 7
-
Check the possible causes and find out the solution!
Possible Causes
What You Can Do?
►
Protein post-translationally modified or alternatively spliced
Check if the protein of interest is post-translationally modified or alternatively spliced and produce other isoforms.
►
Incomplete protein denaturation
Freshly add DTT or β-Mercaptoethanol to sample buffer, or adequaltely boil samples to ensure complete bond breakage between peptides.
►
Membrane protein issue
If a membrane protein is to be detected, try low temperature (~65°C) or avoid boiling which might cause aggregation of membrane proteins.
-
Check the possible causes and find out the solution!
Possible Causes
What You Can Do?
►
Protein post-translationally modified or alternatively spliced
Mix gel completely before pouring. If SDS-PAGE should be run the day after preparation, make sure that the gel is prepared at RT and store in moist chamber at 4°C.
►
Protein degradation
Keep the voltage and temperature low while running SDS-PAGE. If necessary, run gel in the cold room. Remove bubbles trapped at the bottom of gel to ensure even electrophoresis.
►
Protein Multimerization
Make sure that the salt concentration of lysis buffer is kept between 0.15M to 0.5M.
►
Antibody concentration too high
Make sure that total amount of protein loaded into each well is between 20-50μg.
►
Antibody issue
Optimize the dilution factor of primary and secondary antibody.
► Interference from secondary antibody While performing Immunoprecippitaiotn (IP) experiment, make sure that the secondary antibody used to detect the protein of interest is derived from a species different from that of antibody used to pull down protein. Alternatively, use a secondary antibody that recognize only the native form of IgG to detect IP protein.
-
Check the possible causes and find out the solution!
Possible Causes
What You Can Do?
►
Reagent contaminated
Prepare all reagents freshly.
►
Blocking agent insufficiently dissolved
Make sure that the blocking agent such as milk or BSA is completely dissolved before use. Alternatively, filter the blocking solution with 0.45μm filter before use.
-
Check the possible causes and find out the solution!
Possible Causes
What You Can Do?
►
Poor gel preparation
Mix gel completely before pouring. If SDS-PAGE should be run the day after preparation, make sure that the gel is prepared at RT and store in moist chamber at 4°C.
►
Voltage, temperature too high, field effect
Keep the voltage and temperature low while running SDS-PAGE. If necessary, run gel in the cold room. Remove bubbles trapped at the bottom of gel to ensure even electrophoresis.
►
High salt concentration in samples
Make sure that the salt concentration of lysis buffer is kept between 0.15M to 0.5M.
►
Protein overloaded
Make sure that total amount of protein loaded into each well is between 20-50μg.
►
Antibody concentration too high
Optimize the dilution factor of primary and secondary antibody.
-
Check the possible causes and find out the solution!
Possible Causes
What You Can Do?
►
Poor gel preparation
Mix gel completely before pouring. If SDS-PAGE should be run the day after preparation, make sure that the gel is prepared at RT and stored in moist chamber at 4°C.
►
Protein overloaded
Make sure that total amount of protein loaded into each well is between 20-50μg.
►
High membrane protein concentration in samples
If membrane fraction is used, make sure that sample is sufficiently diluted before loading into SDS-PAGE.
ELISA Troubleshooting Guide 5
-
Check the possible causes and find out the solution!
Possible Causes
What You Can Do?
►
Improper standard dilution
Use appropriate diluent as blank. Make sure that the dilution is performed as according to protocol
►
Standard improperly reconstituted
Briefly spin standard vial before opening. Make sure that there is no undissolved material after reconstituting
►
Standard degraded
Store standards as according to protocol
►
Curve doesn’t fit the scale
Try plotting log-log or 5 parameter logistic curve fit
►
Pipetting error
Calibrate pipettes to make sure that the correct volume is dispensed
►
Incomplete washing
Increase washing cycles
-
Check the possible causes and find out the solution!
Possible Causes
What You Can Do?
►
Improper washing
Make sure that the washing is done as according to protocol
►
Poor mixing of samples
Mix samples gently and evenly
►
Dirty plate
Make sure that the bottom of plate is clean
►
Reagents too old
Make sure that the reagents are not expired. Use freshly prepared reagents
►
Bubbles in wells
Make sure that there is no bubble in wells before reading
►
Inconsistent pipetting
Calibrate pipettes to make sure that the correct volume is dispensed
►
Edge effects
Make sure that the plate and reagents are equilibrated to room temperature before starting assay
-
Check the possible causes and find out the solution!
Possible Causes
What You Can Do?
►
Too much antibodies was used
Reduce the concentration of primary or secondary antibodies
►
Antibodies bind nonspecifically
Use blocking buffer or choose another affinity-purified antibody
►
Too much substrate reagent used
Use substrate with higher dilution
►
Insufficient washing
Increase washing cycles
►
Wrong concentration of blocking reagent
Check the recommended concentration of blocking buffer
►
Reaction not stopped
Stop reactions with STOP buffer before reading
►
Plate left too long before reading
Take measurements shortly after addition of substrate and STOP buffer
►
Insufficient Tween in wash buffer
Use PBS+0.05% Tween as wash buffer
►
Incubation temperature too high
Optimize incubation temperature for each experiment
►
Plate stacking during incubation lead to uneven temperature throughout the plate
Avoid stacking plates together during incubation
►
Pipetting error
Calibrate pipettes to make sure that the correct volume is dispensed
►
Reagents not mixed properly
Make sure that all reagents are mixed properly and equilibrated to room temperature before assay
►
Salt concentration of incubation and wash buffer
Increase salt concentration to reduce nonspecific interaction
►
Substrate incubation carried out in light
Perform substrate incubation in dark
►
Dirty plate
Make sure that the bottom of plate is clean
-
Check the possible causes and find out the solution!
Possible Causes
What You Can Do?
►
Insufficient coating
Use more antigens or antibodies for coating
►
Substrate reagents have expired or prepared at a wrong pH
Use fresh substrate reagents
-
Check the possible causes and find out the solution!
Possible Causes
What You Can Do?
►
Assay set up incorrectly
Make sure that the instructions in the protocol is followed carefully
►
Incorrect secondary antibody used
Check if the correct secondary antibody is used
►
Insufficient antibodies used
Increase concentration of primary or secondary antibody
►
Substrate reagents not fresh
Use fresh substrate reagents
►
Wrong settings of plate reader
Check the settings (wavelength, filters, gain etc) of plate reader
►
Insufficient incubation
Follow the incubation time as indicated in the protocol booklet
►
Sample concentration falls below detection limits of kit
Decrease dilution factor or concentrate samples
►
Plate washing too vigorous
Check the setting of plate washer. Pipette wash buffer into wells gently
►
Wells dried out
Cover plate with adhesive film or incubate in humidified chamber throughout experiment
►
Enzyme inhibitor present in buffers or reagents
Inhibitors such as Sodium Azide can affect enzyme and assay performance. Ensure that there is no enzyme inhibitor in any buffers
Immunohistochemistry Troubleshooting Guide 3
-
Check the possible causes and find out the solution!
Possible Causes
What You Can Do?
►
Insufficient blocking
Select appropriate serum as blocking buffer. Blocking for 1 hour at room temperature.
►
Interference from endogenous enzymes
Perform H2O2 or Levamisole quenching.
►
Non-specific binding from primary antibodies
Dilute primary or secondary antibody. Choose another IHC-validated primary antibodies.
►
Inadequate washing
Increase washing cycles/time. Increase salt/detergent concentration
for stronger washes.►
Non-specific binding from secondary antibodies
Perform secondary antibody incubation only. Use pre-adsorbed secondary antibody.
►
Non-specific binding from chromogen
Perform chromogen incubation only. Use other chromogen if necessary.
►
Interference from secondary antibody in multicolor staining
Make sure that the fluorochrome does not overlap with one another.
►
Autofluorescence issue
Make sure that there is no endogenous background caused by tissue itself. Check under fluorescence microscope prior to staining to identify autofluorescence.
-
Check the possible causes and find out the solution!
Possible Causes
What You Can Do?
►
Antigen retrieval too harsh
Optimize retrieval steps to give the best morphology.
►
Tissue sections peeled off slide
Dry samples for 2-4 hours at 60°C.
Tissue with high lipid content (eg breast tissues) should be dried for longer time.►
Sectioning issue
Cut thinner slides for better resolution: 3-5 μm. Use a new/sharper blade.
►
Autolysis has occurred
Fix samples as soon as possible.
Choose other fixatives to accelerate penetration.
Fixative perfusion might be necessary for larger tissues. -
Check the possible causes and find out the solution!
Possible Causes
What You Can Do?
►
Over fixation
Reduce fixation time. Perform antien retrieval to unmask epitopes.
►
Insufficient fixation
Increase fixation time or try other fixative.
►
Fixation process delayed
Fix immediately as tissue is extracted.
►
Permeabilization issue
For nuclear/cytoplasmic proteins, add permeabilization agent (eg Triton X-100, Saponin) in blocking and antibody incubation buffer.
For membrane/tight junction proteins, avoid permeabilization agent.►
Primary antibodies not suitable for IHC
Choose an IHC-validated primary antibodies.
►
Wrong secondary antibody used
Make sure that primary and secondary antibodies match one another.
►
Low expression of protein in tissue samples
Use signal amplification methods (eg: HRP Polymer ARG80982, ARG80967, ARG80966)
►
Insufficient deparafinization
Make sure that parafin is removed completely before staining.